Establishing the daily operating reference density levels
WARNING:
The densitometer of the printer is calibrated at installation.
Authorized service personnel should recalibrate the
densitometer annually or after major service or repair.
This procedure enables you to establish the base line values for:
• Low density
• Mid density
• High density
To establish the daily operating levels, proceed as follows:
1.
Press the Key-operator key to enter the Key-operator mode.
2.
Press the down key seven times, followed by the confirm key to select
‘Quality Control’.
5 Restore config.
6 Calibration
7 Service Actions
8 Quality Control
9 Installation
Reachable with Up/Down arrow keys
Visible
A confirmation screen appears:
3.
Do one of the following:
• Press the Confirm key (YES) to start the QC procedure and proceed
with step 4.
• Press the Escape key (NO) to quit.
4.
Define the tray for printing.
The ‘Select tray’ screen appears:
5.
Press the Up/Down arrow keys to select the desired tray, followed by the
Confirm key.
The following screens appear successively:
Drystar AXYS | Advanced Operation (key-operator mode) |
97
2852E EN 20210601 1655
Содержание DRYSTAR AXYS
Страница 1: ...Drystar AXYS 5367 100 User Manual 2852E EN 20210601 1655 ...
Страница 57: ... The display Drystar AXYS Introduction 57 2852E EN 20210601 1655 ...
Страница 89: ...WARNING Never reuse a jammed film Drystar AXYS Basic Operation operator mode 89 2852E EN 20210601 1655 ...
Страница 128: ...5 Press the OK button on the touchscreen 128 Drystar AXYS Preventive maintenance schedule 2852E EN 20210601 1655 ...
Страница 134: ... Cables transducers and accessories 134 Drystar AXYS Remarks for HF emission and immunity 2852E EN 20210601 1655 ...
Страница 152: ...4 18 2832E EN 20151008 5 Remove plastic bag 152 Drystar AXYS Plug Play Installation Manual 2852E EN 20210601 1655 ...
Страница 161: ...Drystar AXYS Plug Play Installation Manual 161 2852E EN 20210601 1655 ...