5.4.1
Mammography specific Image Adjustments
NOTE:
By default the Mammography printing functionality of the printer is switched off.
Two conditions have to be fulfilled to allow printing of Mammography images:
•
Mammo license is activated. See section 3.6.
•
A DICOM host profile for the Mammo modality with setting "Mammo modality
= yes" has to be created. This 2nd step is described here.
In case the printer has no Mammo modality as input, skip this step.
(1) In the browser pages select "Service Engineer Tools - DICOM profile editor -
New".
(2) Enter Nickname (A).
(3) Enter modality AE title (B).
(4) Select "Mammo Modality = yes" (D).
(5) If a small clear border is required (See Figure 17 next page)
- adjust “image orientation” to “r 180” (C)
- switch on Mammo CWR Rotation* (E).
(6) In case the modality and printer is part of an Agfa Solution, make adjustments as
defined in the solution requirements.
Yes
r180
Nick Name:
Use this profile only for incoming 'Calling AE title':
Mammo modality:
Image orientation:
CR QS Mammo
A
B
D
C
E
QS
OK Cancel Show Defaults Show last Input
HELP
Host Profile: CR QS Mammo
...........................................................
...........................................................
...........................................................
...........................................................
No
Mammo CWR Rotation:
536701gk.cdr
Figure 16
*
CWR =
C
hest
W
all
R
otation
Содержание DRYSTAR AXYS
Страница 7: ......
Страница 98: ......
Страница 106: ......
Страница 250: ......
Страница 290: ......
Страница 314: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 332: ......
Страница 334: ......
Страница 352: ......
Страница 372: ......
Страница 378: ......
Страница 409: ......
Страница 440: ......
Страница 454: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Страница 456: ......