DD+DIS033.07E
Repair and Service
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 3.6 / 22
06-2008 Type
5367/100
Agfa Company Confidential
2.5.3
Resetting the Compliance Film Calibration, Internal Densitometer Calibration and
Thermal Print Head Calibration
NOTE:
Resetting the Compliance Film Calibration, Internal Densitometer Calibration and
Thermal Print Head Calibration is required to be sure, that possibly wrong calibrations
from previous service interventions are deleted.
(1) In the service command session select “1 Show settings – 2 Input Tray”.
(2) Check, which tray is using “DT2M” film material.
(3) In the service command session select “6 Calibration – 6 Compliance film
calibration – 1 Reset compliance film calibration for upper tray or 2 Reset
compliance film calibration for lower tray”. (depends on which film tray is using
DT2M film material).
If both trays contain DT2M film material, select both trays, one after another.
(4) Leave the “service command session”
(5) In the service tools page select “Reset calibration data”
(6) Select “Default CDM” and “Default Profile”. Note, that “Default Profile” button is not
available, if no “Print Head Profile” has been performed in the field yet.
(7) Leave the Reset calibration data” screen.
NOTE:
After this action the printer will show warnings
•
HC302016 Lower tray needs film calibration - or
•
HC301016 Upper tray needs film calibration - and
•
HC306014 Printer needs CDM calibration
This problem will be solved in the next section.
2.5.4
Performing Internal Densitometer Calibration, Film Calibrations and Thermal
Print Head Calibration
NOTE:
The Internal Densitometer Calibration, Film Calibration and Thermal Print Head
Calibration is described in this chapter (3.6 Adjustments and Calibrations) in detail.
Here only a short description is given. For details refer to …
•
2.2 CDM (Internal Densitometer) Calibration
•
•
Содержание DRYSTAR AXYS
Страница 7: ......
Страница 98: ......
Страница 106: ......
Страница 250: ......
Страница 290: ......
Страница 314: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 332: ......
Страница 334: ......
Страница 352: ......
Страница 372: ......
Страница 378: ......
Страница 409: ......
Страница 440: ......
Страница 454: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Страница 456: ......