DD+DIS033.07E
Repair and Service
Electrical and Mechanical Codes, Fuses, LEDs
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
3.3
Power Supply
536634ab.cdr
Figure 3
LED
Status
Color
Function
Overtemp
On
Off
Red
Temperature inside power supply
exceeded permissible limits. Operation not
possible*.
Temperature inside power supply
within permissible limit.
+ 3.3 V
On
Off
Green
+3.3 V present
+3.3 V not present
+ 5 V
On
Off
Green
+5 V present
+5 V not present
+12 V
On
Off
Green
+12 V present
+12 V not present
-12 V
On
Off
Green
-12 V present
-12 V not present
+24 V
On
Off
Green
+24 V present
+24 V not present
13..27 V
On
Off
Green
voltage between 13 V and 27 V
voltage < 13 V
*When the temperature inside the power supply exceeds the permissible limit, printing
can continue for up to 120 seconds. After this delay the power supply switches off the
power and checks the temperature every 20 seconds. As soon as the temperature is
below the permissible limit, it switches on again.
Edition 1, Revision 0
DRYSTAR AXYS
Chapter 3.4 / 9
10-2007 Type
5367/100
Agfa Company Confidential
Содержание DRYSTAR AXYS
Страница 7: ......
Страница 98: ......
Страница 106: ......
Страница 250: ......
Страница 290: ......
Страница 314: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 332: ......
Страница 334: ......
Страница 352: ......
Страница 372: ......
Страница 378: ......
Страница 409: ......
Страница 440: ......
Страница 454: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Страница 456: ......