DD+DIS033.07E
Repair and Service
Electrical and Mechanical Codes, Fuses, LEDs
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 0
DRYSTAR AXYS
Chapter 3.4 / 8
10-2007 Type
5367/100
Agfa Company Confidential
3.2
THANOS Board (plugged onto the DEVIL Board)
LED Color
Function
LD1 yellow
FPGA LED1 - Software controllable
LD2 yellow
FPGA LED1 - Software controllable
LD3 yellow
FPGA LED1 - Software controllable
LD5 green
VCC supply voltage ok
LD6 green
3.3V supply voltage ok
LD8 green
1V8 supply voltage ok
LD9 green
1V8 DDR2 supply voltage ok
LD10 green
VCORE supply voltage ok
LD11 yellow
1V2 supply voltage ok
LD12 yellow
PCI Bus status
LD13 yellow
Fire Wire Link active
LD14 yellow
Fire Wire cable connected
LD15 yellow
CPU Bus active
LD16 red
OFF: FPGA configuration done
ON: FPGA NOT configured => ALARM
LD17 yellow
Software
controllable
LD18 yellow
Software
controllable
LD19 yellow
Software
controllable
LD20 yellow
Software
controllable
Содержание DRYSTAR AXYS
Страница 7: ......
Страница 98: ......
Страница 106: ......
Страница 250: ......
Страница 290: ......
Страница 314: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 332: ......
Страница 334: ......
Страница 352: ......
Страница 372: ......
Страница 378: ......
Страница 409: ......
Страница 440: ......
Страница 454: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Страница 456: ......