Record the respective aim (average) values as the ‘Operating levels’ on Charts
2a and 2b (‘Drystar 5500/5503 Daily Density Control Chart’). Refer to
for mammography QC (optional)’
on page
.
The calculated ‘Operating levels’ should be as follows:
10
These charts will be used for the daily quality test. For more information, refer
to
‘Performing the daily QC test for mammography application (DT 2 Mammo)
on page
Operating Level
Value (Macbeth units)
(according to NEMA standards XR 23-2006)
Base + Fog
0.22 ± 0.03
Low density
0.52 ± 0.07
Mid density
1.35 ± 0.15
High density
2.35 ± 0.15
Содержание Drystar 5500
Страница 1: ...Drystar 5500 5503 Reference manual 2900H EN 20071108 ...
Страница 6: ...6 2900H EN 20071108 DRYSTAR 5500 5503 REFERENCE MANUAL ...
Страница 176: ...176 2900H EN 20071108 Controlling the Drystar 5500 via a remote PC with browser DRYSTAR 5500 5503 REFERENCE MANUAL ...
Страница 226: ...226 2900H EN 20071108 Troubleshooting DRYSTAR 5500 5503 REFERENCE MANUAL ...
Страница 227: ...Equipment information sheet Appendix A ...
Страница 237: ...Drystar media density response data Appendix B ...
Страница 241: ...Quality Control Charts Appendix C ...
Страница 252: ...252 2900H EN 20071108 Quality Control Charts DRYSTAR 5500 5503 REFERENCE MANUAL ...
Страница 253: ...Remarks for HF emission and immunity Appendix D ...
Страница 259: ...Index Chapter E ...
Страница 265: ...265 2900H EN 20071108 Index DRYSTAR 5500 5503 REFERENCE MANUAL ...
Страница 266: ...0413 Printed in Belgium Published by Agfa HealthCare N V B 2640 Mortsel Belgium 2900H EN 20071108 ...