153
2900H EN 20071108
Advanced operation (Key-operator mode)
D
RYSTAR
5500/5503 R
EFERENCE
MANUAL
Quality control for mammography application
(DT 2 Mammo) (optional)
In order to establish and maintain consistent image quality, a regular
evaluation of the image quality is advised.
The Drystar 5500 contains an automatic QC procedure for general
radiography applications (refer to
‘Quality control for general radiography
applications (DT 2 B & DT 2 C)’
on page
) and has been designed to comply
with the grayscale reproduction constancy test, according to the international
standard IEC 1223-2-4. If the option for the mammography application is
activated, another QC procedure is available for the mammography
application. This QC procedure has been designed to comply with the
Mammography Quality Standards Act (MQSA) of the FDA (refer to
on page
). In this chapter we will discuss how to control the
quality of general radiography applications.
Local Regulations may require other procedures.
The Drystar 5500 Hard Copy quality control procedures consist of two main
steps:
•
Before initial use, establishing a number of reference values that will be
used for further follow-up and verifying initial image quality.
•
After establishing these values, performing regular daily, weekly and
annual quality tests.
The results of these tests are recorded on Quality Control Charts.
The QC image (Refer to the
on page
) has
several additional fields where the QC data can be filled in. This image should
be filed as part of the QC procedure.
For more information, please refer to
Repeat the two main steps of the Hard Copy quality procedures for each input tray
loaded with DT 2 Mammo films.
Содержание Drystar 5500
Страница 1: ...Drystar 5500 5503 Reference manual 2900H EN 20071108 ...
Страница 6: ...6 2900H EN 20071108 DRYSTAR 5500 5503 REFERENCE MANUAL ...
Страница 176: ...176 2900H EN 20071108 Controlling the Drystar 5500 via a remote PC with browser DRYSTAR 5500 5503 REFERENCE MANUAL ...
Страница 226: ...226 2900H EN 20071108 Troubleshooting DRYSTAR 5500 5503 REFERENCE MANUAL ...
Страница 227: ...Equipment information sheet Appendix A ...
Страница 237: ...Drystar media density response data Appendix B ...
Страница 241: ...Quality Control Charts Appendix C ...
Страница 252: ...252 2900H EN 20071108 Quality Control Charts DRYSTAR 5500 5503 REFERENCE MANUAL ...
Страница 253: ...Remarks for HF emission and immunity Appendix D ...
Страница 259: ...Index Chapter E ...
Страница 265: ...265 2900H EN 20071108 Index DRYSTAR 5500 5503 REFERENCE MANUAL ...
Страница 266: ...0413 Printed in Belgium Published by Agfa HealthCare N V B 2640 Mortsel Belgium 2900H EN 20071108 ...