DD+DIS155.11E
Repair and Service
Software Releases and Patches
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
CR 10-X / CR Reader / CR 12-X / CR Advanced ReaderChapter 3.8 / Page 5 of 13
04-2013
Type 5151 / 100 / 110 / 200 / 210
Agfa Company Confidential
2
Software Release ARC_1201
Release Date
05/2012
Digitizer SNs
Software ARC_1201 is installed in digitizers with SN
≥
1039 and additionally in SN 1014, 1016, 1018, 1019,
1020, 1028 ex factory.
Availability
Agfa HealthCare Library*
USB flash drive, delivered with the digitizer
Prerequisites
The NX workstation connected to the digitizer requires
software
≥
NX 2.0.8700 / 3.0.8700.
For additional system component compatibility
information refer to following documents:
Solution Structure Data Document, Document ID*
30781759 (
CR/DR Interoperability Matrix, Document ID*
31333326 (
Reference Document
DD+DIS033.12E, Service Bulletin No. 01,
Digitizer CR 10-X released for controlled Delivery
(
* On the Agfa HealthCare Library only the latest software release can be downloaded.
2.1
New Features and solved Issues of Software ARC_1201
A new Error Viewer version 2.0.9.0 is released together with software ARC_1201.
New features of the Error Viewer:
o
Connection between digitizer and PC is checked periodically: Status indicator
switches to "red triple flashing" if a "ping" to the configured PC fails and to
"fast red blinking" if the connection to the Error Viewer fails. The network
connection failure has the higher priority.
o
Title of the Error Viewer panel is changed to "AGFA Digitizer Remote Display".
o
If an error is displayed, the user can browse through all previous errors i.e.
errors messages which were closed by pressing the "OK" button.
o
Displays now the status “connected” or “offline” in the new status bar.
o
The digitizer checks now the version of the Error Viewer: If this version is not
compatible to the digitizer software or has severe known bugs, the digitizer
switches the status indicator to "red fast blinking" to indicate that the
ErrorViewer must be updated to the version which is integrated in the digitizer
software.
Reliability of the barcode reading improved.
Содержание CR 10-X
Страница 8: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers...
Страница 84: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers...
Страница 104: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers...
Страница 116: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers...
Страница 148: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers...
Страница 202: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers...
Страница 228: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers...
Страница 236: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers...
Страница 283: ...1st Edition CR 10 X CR 12 X Type 5151 100 110 200 210 Order no DD DIS155 11E...