Test Phantom
1.
Positioning template orientation label.
2.
Red dot.
Cassette / Phanthom
POS
ITIO
NIN
G
TEM
PLA
TE
AGF
A
AU
TO
QC
2
PH
AN
TO
M
AG
FA
11
22
Figure 10: The Auto QC² test phantom.
The phantom is used for the test exposures PHAP and PHQC. Materials and
elements in the phantom influence exposure result on the cassette. These
results are used to perform calculations on quality and continuity of the X-Ray
equipment and the Digitizer.
Find below a functional and technical description of the phantom:
1.
Step wedge.
2.
Copper plate.
3.
MTF Edge targets.
4.
Horizontal ruler.
5.
Vertical ruler.
6.
ID Data.
7.
Jitter Slits.
8.
Pixel Size Square.
Auto QC² | Auto QC² Hardware Components |
31
2370C EN 20180330 1524