Equipment Classification
Per EN/IEC60601-1, Medical Electrical Equipment, General Requirements for
Safety 3rd Edition, this device is classified as following:
Table 27: Equipment classification
Class I equip-
ment
Equipment in which protection against electric shock does
not relay on basic insulation only, but includes a power sup-
ply cord with protective earth conductor. For earth reliability
always plug the main power cord into an earthed mains
power outlet.
Type B equip-
ment
A Type B applied part is one that provides a particular de-
gree of protection against electric shock particularly regard-
ing allowable leakage current and reliability of the protec-
tive earth protection.
Water ingress IPX0
This device does not have protection against ingress of wa-
ter.
Cleaning
See section on cleaning and disinfecting.
Disinfection
See section on cleaning and disinfecting.
Flammable
anesthetics
The equipment is not type AP or APG.
This device is not suitable for use in the presence of a flam-
mable anesthetic mixture with air, or in presence of a flam-
mable anesthetic mixture with oxygen or nitrous oxide.
Operation
Continuous operation with intermittent load.
DR 100s | Product Information |
165
0411D EN 20220627 1040