Error messages
Details on the contents of messages can be found in the service
documentation which is available to service personnel.
The device must be restarted in order to continue working.
Message
Description
Action
FAULTY POWER
Error on the Charger or
Chopper. No energy
available.
Switch off, wait some
minutes, switch on, if
the error appears gain
call Service.
ERROR V3
Absent V3 power sup-
ply.
Switch off, wait some
minutes, switch on, if
the error appears gain
call Service.
FILAMENT
No filament current.
Switch off, wait some
minutes, switch on, if
the error appears gain
call Service.
ERROR V2
Absent V2 power sup-
ply.
Switch off, wait some
minutes, switch on, if
the error appears gain
call Service.
LOCKED STARTER
Error during the start-
ing up.
Press the reset button
and repeat the expo-
sure.
FAULTY CHOPPER
Error in the chopper
Press the reset button
and repeat the expo-
sure.
X-RAY ERROR
The kV value has not
reached 75% of the re-
quested value within
the first 10ms of expo-
sure.
Press the reset button
and repeat the expo-
sure.
MAX. TIME
The maximum exposure
time has been reached.
Press the reset button
and repeat the expo-
sure.
DATA ERROR
Memory error, data
checksum error.
Switch off, wait some
minutes, switch on, if
the error appears gain
call Service.