Warning messages
Warning messages must be reset using the reset button in order to continue
working.
Figure 13: Reset button
Message
Description
Action
CLOCK ERR.
System clock error
Press the reset button to
continue.
INI. APR
APR checksum error
Press the reset button to
continue.
ERROR IN APR
An APR value is out of
range
Change the value.
TUBE SEASONING
After a long idle period
(3 months or more) it is
necessary to perform X-
ray tube seasoning, to
avoid damage.
Press the reset button to
continue. Call Service
for the tube seasoning.
HOT TUBE
The temperature of the
X-ray tube has reached
the maximum allowed
value.
Wait for the monobloc
to cool down.
EXPIRED TIME
The exposure button
has been pressed half-
way for preparing the X-
ray tube for more than
15 seconds.
Release the exposure
button and repeat the X-
ray exposure sequence.
MANUAL STOP
The exposure button
has been released be-
fore the end of the ex-
posure.
Press the reset button to
continue.
INACTIVE DAP
Dosimeter not connec-
ted.
-