DR workflow for dynamic images
This workflow is available only on DR systems that support dynamic imaging.
To acquire a set of fluo sequences, rapid sequences and static images for
diagnosis:
1.
Add a dynamic group to the
Image Overview
pane.
If a dynamic group has already been added based on data from the RIS,
this step can be skipped.
a) In the
Examination
window, click
Add Image
.
The
Add Image
window appears.
Figure 75: Add Image
b) Specify the exam group and exam type by clicking on the buttons.
c) Select an exam type that is configured as a dynamic group and click
OK
.
The dynamic group thumbnail is added to the
Image Overview
pane.
A dynamic group thumbnail is indicated with an icon in the top right
corner of the thumbnail.
Figure 76: Thumbnail for a dynamic group
2.
Select the thumbnail for the dynamic group in the
Image Overview
pane
of the
Acquisition
window.
The selected DR detector is activated. The default X-ray exposure
parameters and X-ray system position for the selected examination are
sent to the modality. The software console displays these settings in the
examination overview.
DR 800 | Basic Workflow |
117
0392C EN 20210309 1049
Содержание 6010/200
Страница 1: ...DR 800 6010 200 User Manual 0392C EN 20210309 1049 ...
Страница 95: ...DR 800 Safety Directions 95 0392C EN 20210309 1049 ...
Страница 116: ...Resetting the fluo timer on page 203 116 DR 800 Basic Workflow 0392C EN 20210309 1049 ...
Страница 123: ...DR 800 Basic Workflow 123 0392C EN 20210309 1049 ...
Страница 217: ...Patient size mAs Extra Large normal mAs 4 DR 800 Software Console 217 0392C EN 20210309 1049 ...