Equipment Classification
Per EN/IEC 60601-1, EN/IEC 60601-2-54, this device is classified as
following:
Table 32: Equipment classification
Class I equip-
ment
Equipment in which protection against electric shock
does not rely on basic insulation only, but includes a
fixed connection to mains power with protective earth
conductor.
Type B applied
part
A Type B applied part is one that provides a particular de-
gree of protection against electric shock particularly re-
garding allowable leakage current and reliability of the
protective earth protection.
Protection
against ingress of
solid foreign ob-
jects and water
IP10
This device is protected against solid objects with a size
(diameter) of 50 mm or larger.
This device is not protected against droplets of water.
Cleaning
See section on cleaning and disinfecting.
Disinfection
See section on cleaning and disinfecting.
Flammable anes-
thetics
This device is not suitable for use in the presence of a
flammable anesthetic mixture with air, or in presence of
a flammable anesthetic mixture with oxygen or nitrous
oxide.
Operation
Continuous operation.