Positioning the floating tabletop
To release the brake for moving the floating tabletop, press and hold the
middle foot pedal. The tabletop can be moved in longitudinal and transversal
direction manually.
At the center position, movement is stopped. Press and hold the middle foot
pedal again to release the brake and continue moving the floating tabletop.
To stop movement and activate the brake, release the foot pedal.
Figure 68: Foot pedal to release the brake for the floating table top
Have the patient get on or off the table in the center of the table. If the table
top is extended to the maximum length at the head or foot end, the patient
must not sit on the end of the table top, since the weight load can lead to table
deformations and damage to the product.
Figure 69: Getting on and off the radiographic table
In case of very heavy patients the table top has to be positioned in center
before the patient getting on. The table top must remain in center also during
examination.
The radiographic table is designed for a maximum patient
weight of 320 kg. Don’t apply a load of more than 100 kg at the
end of the table top.