Stopping DX-D 100
To stop DX-D 100:
1.
Stop NX and shut down the PC from the Windows Start menu or login
window.
2.
Put the unit in parking position.
a) Use the tube-collimator handles to restore the default position of the
collimator and the X-ray tube.
b) Press and hold the brake control while using the handles to fully
retract the telescopic arm and turn the column until the parking detent
is aligned with the catch.
c) Lower the arm and fully insert the parking detent into the catch.
WARNING:
Lowering the arm when the telescopic arm is not fully
retracted, can trigger the handle bar and cause unintended
displacement of the equipment. Fully retract the telescopic
arm before lowering the arm.
3.
Lock the device.
• To lock the device using the key, switch the key on the control panel to
the “OFF” position.
1
1.
“OFF” position
• To lock the device using the on/off keypad, press and hold the power
button until the green status indicator light goes out.
1
2
1.
Power button
2.
Enter button
Two minutes remain before shutdown. The X-ray generator and the DR
Detector are powered off.
4.
In a configuration with wireless DR Detector, power off the DR Detector:
a) turn off the DR Detector,
68
| DX-D 100 | Getting started