Charts for general radiography QC
Determination of Operating Levels
Imager Type: __________
Serial #: _________________
Date ____________________
Film Type: ____________
Emulsion #: ______________
Densitometer: ________________ (default selection)
Day 1
Day 2
Day 3
Day 4
Day 5
Month
Day
Initials
Low Density
Average of 5 Values = calculated reference L ow De nsi ty l evel
Mid Density
Average of 5 Values = calculated reference "Mid Density" level
High Density
Average of 5 Values = calculated reference Hi gh De nsi ty l evel
Step 1:
Print QC Test images on five consecutive days. Record the optical densities
measurements in the tables below. After five days, average the values to determine
the reference levels for each of the parameters.
Step 2:
Copy the calculated reference levels to Charts 2A/B (’Daily Density Control Chart’)
General radiography applications
Input Tray: _______________
Quality Control for
Chart 1
Figure 10: Chart 1, Determination of Operating Levels
130
| Drystar 5500, Drystar 5503 | Quality Control Charts
2901J EN 20191018 0826
Содержание 5364/100
Страница 1: ...Drystar 5500 Drystar 5503 5364 100 5364 300 User Manual 2901J EN 20191018 0826 ...
Страница 12: ...12 Drystar 5500 Drystar 5503 Drystar 5500 Drystar 5503 2901J EN 20191018 0826 ...
Страница 58: ... Audio signals The keypad The display 58 Drystar 5500 Drystar 5503 Introduction 2901J EN 20191018 0826 ...
Страница 87: ...The menu structure Drystar 5500 Drystar 5503 Advanced Operation key operator mode 87 2901J EN 20191018 0826 ...
Страница 160: ...4 Open the upper lower input tray 160 Drystar 5500 Drystar 5503 Plug Play Installation Manual 2901J EN 20191018 0826 ...