Removing a Jammed Image Plate
The user is allowed to open the front cover, e.g. to solve image plate jams at
the front side. When opening the front panel, all motor driven system
movements will be stopped (including the laser).
To open the front cover:
1
Simultaneously press the two buttons positioned underneath the feed table.
Note: The technical concept does not allow the user to remove the top cover.
The concept provides maximum reliability that no image plate jam can occur
in the post scan area.
Note: The digitizer always reads and digitizes the plate first, then erases it
and transports it back into the cassette. If a plate jam occurs before the plate
is scanned, there is a fair chance that you can recover the image by putting
the image plate back into the cassette and digitizing it again. While handling
the image plate, prevent exposing it to daylight as much as possible.