Introduction
CR 30-X/CR 30-X
M
Connectivity
The digitizer is connected to the workstation via Ethernet connection and uses
the DICOM protocol to communicate with the workstation.
Installation
The digitizer is equipped with 2 handles at the bottom left and right sides to
move the device easily to another location. It is recommended to have at least
two persons lift the digitizer.
Caution:
When installing the digitizer, care must be taken to ensure that
there is either a mains plug or an all-cable disconnecting device in the
internal installation fitted near the digitizer and that it is easily
accessible.
Caution:
The digitizer and the cassette storage shall be protected against
direct radiation in such a way, that the annual dose equivalent at the
place of installation will not exceed 1 mSv/a.
Caution:
Do not lift the device by holding the input tray.
Caution:
If the digitizer is installed inside an X-ray room it must be
protected from stray radiation by proper shielding.