Chapter 2 Dispensing to SpectroCHIP Arrays
43
USG-CUS-059 Rev05
MassARRAY® Nanodispenser RS1000 v2.1 User Guide
For Research Use Only. Not for use in diagnostic procedures.
2.7 Setting up the instrument deck
Filling the calibrant
reservoir
NOTE
NOTE
If you are not dispensing calibrant, skip this section and proceed to
Loading the microtiter plates,
The calibrant reservoir is a single-well insert located on the dry station. The instrument
aspirates calibrant from this reservoir.
To fill the calibrant reservoir:
1.
Allow the iPLEX or MassCLEAVE
calibrant to equilibrate to room temperature for 5
minutes (if it has been refrigerated) or 10 minutes (if it has been frozen).
2.
Tap the
PARK
button
to
park the pin array assembly.
The pin array assembly moves to allow access to the processing deck.
3.
Open the main door.
The
safety interlock is disengaged! message appears.
4.
Fill the calibrant reservoir with 60
L of calibrant using a pipette.
NOTE
NOTE
When filling the calibrant reservoir, be sure to insert the pipette tip to the bottom of the reservoir
well. Otherwise, some of the calibrant may adhere to the sides of the reservoir rather than collect
at the bottom of the well.
Figure 2.35 Calibrant Reservoir
Calibrant
reservoir
Add calibrant
to this well
Dry station