OM120N-I Operating instructions:     
 

Control Key’s: (All key’s are rubber key pads)  

1.

 

On\Off  

2.

 

Backlight 

3.

 

Wavelength Select (850nm, 1300nm, 1310nm, 1490nm, 1550nm & 1625nm) 

4.

 

dB\dBm with mW Displayed 

5.

 

Set Resolution in dBm mode or Zero reference in dB mode 

 
Backlight Key allows the user the option to turn the display backlight on. 
 
The Wavelength select key allows the user to select between the calibrated wavelengths 
of the OM120N-I. All wavelengths are calibrated to +/- 0.3 dB of the NIST standard. 
 
The  dB\dBm  select  key  allows  the  user  to  select  between  absolute  power,  which  is 
measured  in  dBm,  or  relative  power  that  is  measured  is  dB.  dBm  is  referenced  to  1 
milliwatt and both are a function of the log scale. 
 
The  Rez  key  when  in  the  dBm  mode  allows  the  users  to  toggle  the  resolution  between 
0.01dB and 0.1dB. 
 
The  Zero  reference  key  when  in  the  dB  mode  allows  the  user  to  set  the  input  launch 
power to a relative dB zero reference value for easy calculation of loss and attenuation of 
a fiber optic cable. For instance, if a launch cable output power value is –5.6 dBm and a 
test cable is inserted between the launch cable and the meter with a reading of –7.1 dBm, 
the user has to subtract –7.1 dBm from –5.6 dBm to get a result of  –1.5 dB. As simple as 
this test sounds it can sometimes get confusing to the user due to the fact that they are not 
sure if their result should be a positive value or a negative value also miscalculation can 
be an issue. The zero reference function eliminates these problems. The user simply sets 
their reference value (which in this case is –5.6 dBm) to 0.0 dB then inserts the test cable 
to get a loss reading of  –1.5 dB. *EASY*  
Different  references  can  be  stored  for  all  four  wavelengths.  The  stored  reference  value 
will be maintained until a new reference is stored. 
The  screen  also  displays  the  value  that  the  user  has  referenced  too  at  the  bottom  left 
corner  of  the  screen.  This  reading  is  the  dBm  (absolute)  value  of  the  users  dB  zero 
reference value.    
 
* The OM120N-I also offers an auto power down feature. This is selectable on power up. 
User  can  select  one  of  three  options 

  1)  no  Auto  Power  off      2)AO10  –  power  down 

after 10 minutes 3) AO30 – power down after 30 minutes.  

 

   

Advanced Fiber Solutions, Inc.  

              

 

     19 Norfolk Ave, Easton, Ma 02375, USA 

Fax: (617) 507-0784  

 

 

 

 

     email:[email protected] 

 

Ph: (508) 238-7100  

 

 

 

 

     url:www.afsi.us 

 
 

Product:

 

AF-OM120N-I 

Date: Mar:2013 