2.2.1. AC Power Connection
AC input power to the Ensemble LAB is applied to the AC power receptacle that is located on the rear panel.
The power cord connected to this receptacle also provides the protective earth ground connection and may
serve as a Mains disconnect. The main power switch is located on the front panel of the Ensemble LAB.
Refer to
for the electrical specifications.
W A R N I N G : The AC power cord is the Mains disconnect.
Table 2-1:
Main AC Power Input Voltages and Current Requirements
AC Input Voltage
Input Amps (maximum continuous)
Connector Port
100 VAC 50/60 Hz
2 A
115 VAC 50/60 Hz
2 A
200 VAC 50/60 Hz
1 A
230 VAC 50/60 Hz
1 A
The AC power cord to the Ensemble LAB must be rated for the line voltage being used and have a minimum
current capacity of 5 A. The insulation for the AC power wiring must be appropriately rated for the
environment. The temperature rating of the insulation must be at least 80ºC. Environmental conditions may
necessitate the need to meet additional AC wiring requirements or specifications. AC wiring should not be
bundled with signal wiring to minimize EMI coupling and interference.
D A N G E R : See the user documentation provided with your Ensemble LAB system to
determine if the Ensemble LAB chassis is limited to only one AC input voltage. Operation at
other voltages may result in damage to the Ensemble LAB chassis.
www.aerotech.com
Chapter 2
23