Page | 38
Aeroqual AQM60 User Guide
To remove a run, right click inside the run scheduler and select remove run. Note that run three will be
removed first followed by run two and then run one. Once a run has been removed click save and then click
read to make sure that the run has been removed.
Note: The information contained in the Calibration Configuration is stored on the AQM controller and
not on the PC containing the Aeroqual software.
4.5.5. The Calibration Record
Each time a scheduled calibration run is executed a calibration log (CL) is written to the s.d card in the date
time format: CLMMDDHH, where MM = month, DD = day of month, HH = hour. An example of the
calibration log is shown below.
Monitor ID: 1
Port Specification
Port
Gas
Name
Conc/
ppm
Gas
Name
Conc/
ppm
Gas
Name
Conc/
ppm
1 Isobu
100
0
0
2 CO
250
0
0
Day Time
Port
CO
/ppm
PID
/ppm
MFM Z
/ml/min
MFC S
/ml/min
TEMP
/C
RH
/%
DR
5/10/2011 14:39
1
0.5
7.89
1129.2
122
25.96
44.8
10
5/10/2011 14:41
1
0.24
8.36
1137.5
119
25.96
44.8
10
5/10/2011 14:43
1
0.34
9.65
1134
126
26
44.6
10
5/10/2011 14:45
1
0.33
9.97
1143.8
127
26
44.4
10
5/10/2011 14:47
1
0.29
10.01
1131.1
128
25.95
44.5
10
5/10/2011 14:49
1
0.24
10.03
1130.7
125
25.97
44.5
10
5/10/2011 14:51
1
0.18
10.12
1130.6
125
26.03
44.5
10
5/10/2011 14:53
1
0.13
10.08
1131.1
123
26.01
44.5
10
5/10/2011 14:55
1
0.07
10.11
1121.8
122
26.04
44.5
10
5/10/2011 14:57
1
0.01
10.1
1144.3
126
26.14
44.3
10
5/10/2011 14:59
1
0.02
10.07
1142.1
126
26.19
44.1
10
5/10/2011 15:01
1
0.08
10.09
1130.6
125
26.19
44
10
5/10/2011 15:03
1
0.12
10.11
1125.7
124
26.15
44.1
10