A310B Fingertip Pulse Oximeter 

 

Operator’s Manual 

 

      

     

            

 

 

 

 

 

File No.: A310B-OPE        V 1.0      20180712     

Содержание A310B

Страница 1: ...A310B Fingertip Pulse Oximeter Operator s Manual File No A310B OPE V 1 0 20180712...

Страница 2: ...every 2 hours SpO2 measurements may be adversely affected in the presence of high ambient light Shield the sensor area with a surgical towel or direct sunlight for example if necessary The following...

Страница 3: ...to explode CAUTION Keep the operating environment free of dust vibrations corrosive or flammable materials and extremes of temperature and humidity CAUTION Do not operate the unit if it is damp or wet...

Страница 4: ...used if the battery cover is closed CAUTION The batteries must be proper disposed according to local regulation after their use CAUTION The batteries must be taken out from the battery compartment if...

Страница 5: ...Equipment Batch code Date of manufacture SN Serial NO Information of manufacture including name and address Temperature limitation Bluetooth Indication When the end user wishes to discard this product...

Страница 6: ...rmation you should know to protect the equipment from possible damage Note The important information you should know Batch code Date of manufacturer and Serial No are printed on the label on the batte...

Страница 7: ...use in internist surgery Anesthesia pediatrics intensive care and etc Oxygen Bar social medical organizations physical care in sports and etc 2 3 Product Features Lightweight for carrying and Easy To...

Страница 8: ...measured arterial hemoglobin saturation value SpO2 of the sensor is compared to arterial hemoglobin oxygen SaO2 value determined from blood samples with a laboratory CO oximeter The accuracy of the s...

Страница 9: ...invasive saturation of pulse blood oxygen is low accurate Plethysmograph is an instrument for measuring changes in volume within an organ or whole body usually resulting from fluctuations in the amoun...

Страница 10: ...em Name Description 1 Power button Turn on the machine 2 OLED Panel Display the SPO2 PR data Plethysmogram etc 3 Battery Compartment 3 2 Display After switch on the OLED display of A310B is as follows...

Страница 11: ...the setting through moving the symbol to the back of the Sound Reminder Beep Bluetooth Restore or Brightness Sound Reminder Press the direction button for 1 second move the symbol to the back of Sound...

Страница 12: ...User can press the direction button to select the items And press the direction button for 1 second to change the data you need On the Reminder Limit setup menu page figure 3 3 2 when the symbol shown...

Страница 13: ...customer can preset the alarm value to the 98 or 99 to check whether it is normal for alarm setting 3 4 Operation 3 4 1 Install battery Installing two AAA batteries into battery cassette in correct po...

Страница 14: ...Press power button to turn the Pulse Oximeter on The oximeter will be automatically powered off when no finger in the device for longer than 16 seconds 3 4 3 Read correspondent data from display scre...

Страница 15: ...an been regarded as correct if the wave is fluctuated regularly 3 4 5 Date Transmission 1 The current measurement will transmit to the App automatically after data transmission successfully the measur...

Страница 16: ...he power and take out the batteries before cleaning Cleaning exterior surface OLED display screen included of the unit with a dry and soft cloth Use 75 density of medical alcohol to clean the surface...

Страница 17: ...ambient temperature is 10 50 and humidity is 15 80 Regular inspection to make sure that no obvious damage existed to affect the safety and performance of device No flammable substance overtop or lowe...

Страница 18: ...if you can make sure about no problem existing in the product Please go to a hospital timely for exact diagnosis Oxyhemoglobin or heart rate is shown unstably 1 Finger might not be plugged deep enoug...

Страница 19: ...batteries 3 Please contact with local customer service center he screen are suddenly off 1 The product is automatically powered off when no signal is detected longer than 8 seconds 2 Power quantity of...

Страница 20: ...ernally powered equipment Anti electric Shock Degree Type BF equipment EMC Type B Mode of operation Continuous Operation Enclosure Degree of ingress protection IP22 IP22 means shell of this product ca...

Страница 21: ...non condensing Air Pressure 86Kpa 106Kpa Alarm default value Parameter Value Hemoglobin saturation Upper limit 100 bottom limit 94 Pulse rate Upper limit 130 bottom limit 50 Electronics Parameters Par...

Страница 22: ...Accuracy Hemoglobin Saturation 3 70 100 Unspecified 70 Pulse rate 1 BPM PI Display 0 20 Resolution 0 1 Measure Accuracy 0 1 0 1 1 20 1 Probe LED Specification Wave Length Radiant Power RED 660 2 nm 1...

Страница 23: ...adiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference w...

Страница 24: ...may not cause harmful interference 2 this device must accept any interference received including interference that may cause undesired operation 15 21 Information to user Any Changes or modifications...

Страница 25: ...ith RF exposure requirements a minimum separation distance of 5mm for body worn must be maintained between the user s body including the antenna Third party belt clips holsters and similar accessories...