E N G L I S H

User Manual

  VOLTAGE AND CURRENT DATA LOGGERS

Simple Logger

®

 II

Series

Содержание Simple Logger II L101

Страница 1: ...E N G L I S H User Manual VOLTAGE AND CURRENT DATA LOGGERS Simple Logger II Series...

Страница 2: ...ir and calibration facility for a nominal charge The recommended calibration interval for this instrument is 12 months and begins on the date of receipt by the customer For recalibration please use ou...

Страница 3: ...2 2 Control Features 8 3 SPECIFICATIONS 10 4 OPERATION 17 4 1 LED Control Operation and Status Function 17 4 2 Connecting the Simple Logger II to a Computer 19 4 3 Turning the Unit On 19 4 4 Recordin...

Страница 4: ...APPENDIX A TROUBLESHOOTING 28 APPENDIX B GLOSSARY 29 Repair and Calibration 30 Technical and Sales Assistance 30 Limited Warranty 31 Warranty Repairs 31...

Страница 5: ...ent is protected by double or reinforced insulation Indicates aWARNING and that the operator must refer to the user manual for instructions before operating the instrument Failure to follow or carry o...

Страница 6: ...g your shipment make sure that the contents are consistent with the packing list Notify your distributor of any missing items If the equipment appears to be damaged file a claim immediately with the c...

Страница 7: ...ser manual and warranty card Simple Logger II Model L562 Cat 2126 35 2 Channel TRMS Voltage Current Module Includes USB cable DataView CD ROM 2x1 5V AA cell alkaline batteries one set of leads one set...

Страница 8: ...lor coded 5 ft w color coded alligator clips Rated 600V Cat IV 15A Cat 2140 62 Lead Set of 2 Color coded 10 ft w color coded alligator clips Rated 600V Cat IV 15A Cat 2140 63 Replacement A to 5 pin mi...

Страница 9: ...ight times per second The measurement process is performed no more than eight times per second TRMS calculations are performed on 64 samples of a single line cycle for AC measurements This means that...

Страница 10: ...One 4 position removable screw type terminal block L562 BNC Two recessed 4mm safety banana jacks L642 Two miniature thermocouple connectors 2 Five LED Indicators The LEDs on the logger serve two funct...

Страница 11: ...nstrument 5 Reset Switch not shown The RESET switch resets the CPU and is located under the top cover To access this switch remove the battery compartment cover then remove the four screws holding the...

Страница 12: ...15 minutes to 8 weeks programmable using DataView Memory 240 000 measurement 512KB Recorded data is stored in non volatile memory and retained even if battery is low or removed Communication USB 2 0 o...

Страница 13: ...ew Memory 240 000 measurement 512KB Recorded data is stored in non volatile memory and retained even if battery is low or removed Communication USB 2 0 optically isolated Power Source 2x1 5V AA LR6 al...

Страница 14: ...Recorded data is stored in non volatile memory and retained even if battery is low or removed Communication USB 2 0 optically isolated Power Source 2x1 5V AA LR6 alkaline batteries Battery Life 100 h...

Страница 15: ...ng Length 15 minutes to 8 weeks programmable using DataView Memory 240 000 measurement 512KB Recorded data is stored in non volatile memory and retained even if battery is low or removed Communication...

Страница 16: ...top FIFO and Extended Recording Mode XRM Recording Length 15 minutes to 8 weeks programmable using DataView Memory 240 000 measurements 512KB Recorded data is stored in non volatile memory and retaine...

Страница 17: ...Maximum of 8 samples taken at storage interval Storage Rate Programmable from 5 sec to 1 day Storage Modes Start Stop FIFO and Extended Recording Mode XRM Recording Length 15 minutes to 8 weeks progra...

Страница 18: ...RONMENTAL Operating Temperature 14 to 122 F 10 to 50 C Storage Temperature 4 to 140 F 20 to 60 C Relative Humidity Up to 85 95 F 35 C Non condensing Altitude 2000m SAFETY Safety Rating EN61010 1 600V...

Страница 19: ...the PRESS button while the instrument is ON each LED will light in sequence Continuing to hold the PRESS button until the last LED lights turns the instrument OFF At this point releasing the button wi...

Страница 20: ...ble blink At least one channel is currently in an alarm condition Fast blink Armed to clear alarm indication RED LED CONTROL ERASES THE MEMORY see 4 7 STATUS OFF No data in memory Single blink Memory...

Страница 21: ...USB communication port on your computer Refer to the computer manual to locate the USB port on your computer The logger can be connected to the computer during a recording session however an addition...

Страница 22: ...ple Logger II section inside the DataView user manual Once a configuration is written to the instrument the logger will no longer need to be connected to DataView to start the scheduled recording When...

Страница 23: ...urements stored in the instrument are transferred to a computer via the download command in the SLII Control Panel For instructions on downloading data see the Downloading Recorded Instru ment Data se...

Страница 24: ...on is not pressed within five seconds of arming the erase operation will automatically disarm and memory will be maintained For this reason if you do not intend to erase memory simply wait until the R...

Страница 25: ...N LED single blinks STANDBY mode is active and the logger is not recording The RED LED is OFF indicating there is no data in memory The PRESS button is used to Start Stop a Recording Session If the PR...

Страница 26: ...n Start Stop Mode If the logger is recording using the Start Stop mode and memory is filled before the Recording Session has finished the session will end The following happens The GREEN LED single bl...

Страница 27: ...ing in START STOP mode fills the memory The user terminates the session by pressing the PRESS button until the STOP ORANGE LED lights up and releasing the button before the next LED lights or issues a...

Страница 28: ...wer If the battery voltage is below 1 7V the unit will not respond to a button press In this condition pressing the RESET switch will not restart the unit It is recommended to download any desired ses...

Страница 29: ...the PRESS button for 2 seconds to turn ON NOTE If the instrument is stored without the batteries installed the internal clock will need to be reset using the Simple Logger II Control Panel software on...

Страница 30: ...t LED lights Make sure the RED LED is not double blinking If it is memory is full and you need to erase the data see 4 7 Make sure the Simple Logger II is properly configured so that you have Storage...

Страница 31: ...d environment Download The process of transferring data from the logger to the computer Hz Hertz a unit of measure of frequency equivalent to cycles per second I O Input output a device or port capabl...

Страница 32: ...s calibration certificate plus recorded calibration data Ship To Chauvin Arnoux Inc d b a AEMC Instruments 15 Faraday Drive Dover NH 03820 USA Phone 800 945 2362 Ext 360 603 749 6434 Ext 360 Fax 603 7...

Страница 33: ...return the instrument to us for repair provided we have your warranty registration information on file or a proof of purchase AEMC Instruments will at its option repair or replace the faulty material...

Страница 34: ...32 Simple Logger II Series Notes...

Страница 35: ......

Страница 36: ...10 08 99 MAN 100309 v5 Chauvin Arnoux Inc d b a AEMC Instruments 15 Faraday Drive Dover NH 03820 USA Phone 603 749 6434 Fax 603 742 2346 www aemc com...