OPERATION 

Please make sure that you have already read and fully understand the WARNING section on page 1. 

Making Measurements with the AC Current Probe Model

 

MN379 

Connect the black lead of the current probe to “common” and the red lead to the AC voltage input on 
your DMM or other voltage-measuring instrument. The AC current probe has an output of 200mV/A and 
10mV/A. This means that with the probe in the 10mV/A position, for 100A AC in a conductor around 
which the probe is clamped, 1V AC will come out of the probe leads to your DMM or instrument. Select 
the range which corresponds to the measured current. If the current magnitude is unknown, start with 
the  probe  in  the  10mV/A  position  and  if  the  current  is  <  5  amps  the  probe  can  be  switched  to  the 
200mV/A position for the higher resolution. Clamp the probe around the conductor. With the probe in 
the 10mV/A position take the reading on the meter and multiply it by 100 to obtain the measured current 
(e.g, 160mV reading = 160 x 100 = 16,000mA or 16A). 

The 5A range is design to measure on 5A current transformer. The output is 1V rms for 5A measured. 
Enter a scale factor in the meter if possible for direct reading. 

For best accuracy, avoid if possible, the proximity of other conductors which may create noise. 

Tips for Making Precise Measurements 

When  using  a  current  probe  with  a  meter,  it  is  important  to  select  the  range  that  provides  the  best 
resolution. Failure to do this may result in measurement errors. 

Make sure that probe jaw mating surfaces are free of dust and contamination. Contaminants cause air 
gaps between the jaws, increasing the phase shift between primary and secondary. It is very critical for 
power measurement. 

MAINTENANCE: 
Warning

 

For maintenance use only original factory replacement parts. 

To avoid electrical shock, do not attempt to perform any servicing unless you are qualified to do so. 

To avoid electrical shock and/or damage to the instrument, do not get water or other foreign agents 
into the probe 

Cleaning 

To ensure optimum performance, it is important to keep the probe jaw mating surfaces clean at all times. 
Failure to do so may result in error in readings. To clean the probe jaws, use very fine sand paper (fine 
600) to avoid scratching the jaw, then gently clean with a soft oiled cloth. 

REPAIR AND CALIBRATION 

You  must  contact  our  Service  Center  for  a  Customer  Service  Authorization  number  (CSA#).  This  will 
ensure that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA# 
on the outside of the shipping container.  

Chauvin Arnoux

®

, Inc. d.b.a. AEMC

®

 Instruments 

15 Faraday Drive 

 Dover, NH 03820 USA 

(800) 945-2362 (Ext. 360) • (603) 749-6434 (Ext. 360) • [email protected] 

(Or contact your authorized distributor) 

NOTE: All customers must obtain a CSA# before returning any instrument. 

TECHNICAL AND SALES ASSISTANCE 

If  you  are  experiencing  any  technical  problems,  or  require  any  assistance  with  the  proper  use  or 
application of this instrument, please contact our technical hotline: 

(800) 945-2362 (Ext. 351) • (603) 749-6434 (Ext. 351) 

 • [email protected] 

LIMITED WARRANTY 

The  current  probe  is  warranted  to  the  owner  for  a  period  of 

two

  year

s

  from  the  date  of  original  pur

chase against  defects  in  manufacture.  This  limited  warranty  is  given  by  AEMC

®

  Instruments,  not 

by  the distributor from whom it was purchased. This warranty is void if the unit has been tampered with, 
abused or if the defect is related to service not performed by AEMC

®

 Instruments.  

Full warranty coverage and product registration is available on our website at: 
www.aemc.com/warranty.html. 

Please print the online Warranty Coverage Information for your records. 

AC Current Probe 
Model MN379 

User Manual

 

DESCRIPTION 

The MN379 (Catalog #2153.01) is the latest in compact AC current probes. Designed to meet the most 
stringent demands in industry and electrical contracting, it also meets the latest safety and performance 
standards.  The  probe  has  a  measurement  range  up  to  120Arms,  a  range  and  resolution  capable  of 
displaying 10mV of output per amp of measured current, a voltmeter accuracy of 0.5% or better to take 
full advantage of the accuracy of the probe and an input impedance of 



1M

WARNING 

The  safety  warnings  are  provided  to  ensure  the  safety  of  personnel  and  proper  operation  of  the 
instrument. Read the instruction completely. 

•  Use caution on any circuit: potentially high voltages and currents may be present and may pose 

a shock hazard. 

•  Do not use the probe if damaged. Always connect the current probe to the measuring device 

before it is connected around the conductor 

•  Do  not  use  on  non-insulated  conductor  with  a  potential  to  ground  greater  than  600V  CAT  III 

pollution 2. Use extreme caution when clamping around bare conductors or bus bars. 

•  Before each use, inspect the probe; look for cracks in housing or output cable insulation. 

•  Do not use clamp in wet environment or in locations that hazardous gases exist. 

•  Do not use the probe anywhere beyond the tactile barrier. 

INTERNATIONAL ELECTRICAL SYMBOLS 

This symbol signifies that the current probe is protected by double or reinforced insulation. Use 
only factory specified replacement parts when servicing the instrument. 

This symbol signifies CAUTION! and requests that the user refer to the user manual before 
using the instrument. 

This is a type A current sensor. This symbol signifies that application around and removal 
from HAZARDOUS LIVE conductors is permitted. 

DEFINITION OF MEASUREMENT CATEGORIES 

CAT II: 

For  measurements  performed  on  circuits  directly  connected  to  the  electrical  distribution  system. 
Examples are measurements on household appliances or portable tools. 

CAT III:

  For  measurements  performed  in  the  building  installation  at  the  distribution  level  such  as  on 

hardwired equipment in fixed installation and circuit breakers. 

CAT IV: 

For  measurements  performed  at  the  primary  electrical  supply  (<1000V)  such  as  on  primary 
overcurrent protection devices, ripple control units, or meters. 

RECEIVING YOUR SHIPMENT 

Upon receiving your shipment, make sure that the contents are consistent with the packing list. Notify 
your distributor of any missing items. If the equipment appears to be damaged, file a claim immediately 
with the carrier and notify your distributor at once, giving a detailed description of any damage  