background image

99-MAN 100158.v5 09/07 

OPERATION 

Please make sure that you have already read and fully understand the WARNING section on page 1. 

Making Measurements with the AC Current Probe Model MN306/MN307 

  Connect the black lead of the current probe to “common” and the red lead to the AC voltage input on 

your DMM or other voltage-measuring instrument. The AC current probe has an output of 100mV/A. 
This means that for 10A AC in a conductor around which the probe is clamped, 1V AC will come out of 
the probe leads to your DMM or instrument. Select the range which corresponds to the measured 
current. If the current magnitude is unknown, start with the highest range and work down until the 
appropriate range and resolution are reached. Clamp the probe around the conductor. Take the 
reading on the meter and multiply it by 10 to obtain the measured current (e.g, 160mV reading = 160 x 
10 = 1,600mA or 1.6A). 

  For best accuracy, avoid if possible, the proximity of other conductors which may create noise. 

Tips for Making Precise Measurements 

  When using a current probe with a meter, it is important to select the range that provides the best 

resolution. Failure to do this may result in measurement errors. 

  Make sure that probe jaw mating surfaces are free of dust and contamination. Contaminants cause air 

gaps between the jaws, increasing the phase shift between primary and secondary. It is very critical for 
power measurement. 

MAINTENANCE: 

Warning

  For maintenance use only original factory replacement parts. 

  To avoid electrical shock, do not attempt to perform any servicing unless you are qualified to do so. 

  To avoid electrical shock and/or damage to the instrument, do not get water or other foreign agents 

into the probe 

Cleaning 

To ensure optimum performance, it is important to keep the probe jaw mating surfaces clean at all times. 
Failure to do so may result in error in readings. To clean the probe jaws, use very fine sand paper (fine 
600) to avoid scratching the jaw, then gently clean with a soft oiled cloth. 

REPAIR AND CALIBRATION 

You must contact our Service Center for a Customer Service Authorization number (CSA#). This will 
ensure that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA# 
on the outside of the shipping container. If the instrument is returned for calibration, we need to know if you 
want a standard calibration, or a calibration traceable to N.I.S.T. (includes calibration certificate plus 
recorded calibration data).  

Chauvin Arnoux

®

, Inc. d.b.a. AEMC

®

 Instruments 

15 Faraday Drive 

 Dover, NH 03820 USA 

Tel: 

(800) 945-2362  (Ext. 360) 
(603) 749-6434  (Ext. 360) 

 Fax:

(603) 

742-2346 

 [email protected] 

(Or contact your authorized distributor) 
Costs for repair, standard calibration, and calibration traceable to N.I.S.T. are available. 

NOTE: All customers must obtain a CSA# before returning any instrument

TECHNICAL AND SALES ASSISTANCE 

If you are experiencing any technical problems, or require any assistance with the proper use or 
application of this instrument, please call our technical hotline: 

(800) 343-1391 • (508) 698-2115 • Fax (508) 698-2118 

Chauvin Arnoux

®

, Inc. d.b.a. AEMC

®

 Instruments 

[email protected] 
www.aemc.com 

AC Current Probe 
Models MN306 and MN307 

User Manual

 

DESCRIPTION 

The MN306/307 (Catalog #2116.22 and #2116.23) are the latest in compact AC current probes. 
Designed to meet the most stringent demands in industry and electrical contracting, they also meet the 
latest safety and performance standards. The probes have a measurement range up to 12Arms which 
make them perfect tools for measurements with DMMs, recorders, power and harmonic meters. The 
Models MN306/MN307 are compatible with any AC voltmeter, multimeter, or other voltage 
measurement instrument that is capable of displaying 100mV of output per amp of measured current, 
voltmeter accuracy of 0.75% or better and an input impedance of 

 1M

Ω

WARNING 

These safety warnings are provided to ensure the safety of personnel and proper operation of the 
instrument. 

Read the instruction manual completely and follow all the safety information before attempting
to use or service this instrument. 

Use caution on any circuit: Potentially high voltages and currents may be present and may
pose a shock hazard. 

Read the specifications section prior to using the current probe. Never exceed the maximum
voltage ratings given. 

•  Safety is the responsibility of the operator. 

ALWAYS connect the current probe to the display device before clamping the probe onto the
sample being tested. 

ALWAYS inspect the instrument, probe, probe cable, and output terminals prior to use.
Replace any defective parts immediately.

NEVER use the current probe on electrical conductors rated above 600V in overvoltage
category III (CAT III). Use extreme caution when clamping around bare conductors or bus bars. 

INTERNATIONAL ELECTRICAL SYMBOLS 

This symbol signifies that the current probe is protected by double or reinforced insulation. Use 
only factory specified replacement parts when servicing the instrument. 

This symbol signifies CAUTION! and requests that the user refer to the user manual before 
using the instrument. 

This is a type A current sensor. This symbol signifies that application around and removal from 
HAZARDOUS LIVE conductors is permitted. 

DEFINITION OF MEASUREMENT CATEGORIES 

Cat. I:

  For measurements on circuits not directly connected to the AC supply wall outlet such as 

protected secondaries, signal level, and limited energy circuits. 

Cat. II: 

For measurements performed on circuits directly connected to the electrical distribution system. 
Examples are measurements on household appliances or portable tools. 

Cat. III:

 For measurements performed in the building installation at the distribution level such as on 

hardwired equipment in fixed installation and circuit breakers. 

Cat. IV: 

For measurements performed at the primary electrical supply (<1000V) such as on primary 
overcurrent protection devices, ripple control units, or meters. 

RECEIVING YOUR SHIPMENT 

Upon receiving your shipment, make sure that the contents are consistent with the packing list. Notify 
your distributor of any missing items. If the equipment appears to be damaged, file a claim immediately 
with the carrier and notify your distributor at once, giving a detailed description of any damage. 

99 Washington Street 
Melrose, MA 02176 

Phone

 781-665-

1400

Toll Free 1-800-517-8431

Visit us at www.T

estEquipmentDepot.com

Отзывы: