R3765/67G Series Network Analyzer Operation Manual
3.6 Measurement Samples of Simple Reflection Characteristics
3-22
(3)
Setting
Press [PRESET] to preset.
Setting of the R3765/67G series is as follows.
Note:
OPT11 or OPT14 uses a different [MEAS] menu.
For more information, refer to the description on page 7-10.
3.6.2
Calibration (1-port full calibration)
In the case of AG type, perform 1-port full calibration of bridge test port.
In the case of BG/CG type, perform 1-port full calibration of test port 1.
CAUTION!
1. If the calibration has already been executed, switch OFF the calibration, clear the calibration
data, and then start calibration. (Refer to Section "7.5.10 Calibration Data Clear.")
2. When the message “Wait for Sweep” disappeared, each calibration completes.
The above message may not be visible if measurement time is short. A beep sound is always
heard when the calibration operation is complete.
3. During it’s displayed the R3765/67G series, the cable, the connector, etc. must not be moved.
Select the calibration kit to be used. (refer to Section 7.5.7 Calibration Kit Selection).
Invoke 1-port full calibration menu.
[CAL]
{CAL MENU}
{1 PORT FULL CAL}
Block name
Setting
Key operation
ACTIVE
CHANNEL
Set the channel to 1.
[CH 1]
(Initial setup)
RESPONSE
Select the input port in the
receiver part.
Set the measurement
format to magnitude (log
display).
AG type : Bridge is used.
[
MEAS
]
{A/R}
(Initial setup)
BG type :
[MEAS
]
{REFLECTION}
(Initial setup)
CG type :
[MEAS
]
{S11 REFL FWD}
(Initial setup)
[FORMAT
]
{LOG MAG}
(Initial setup)
STIMULUS
Center frequency
880MHz
Span frequency
100MHz
[CENTER]
[8]
[8]
[0]
[MHz]
[SPAN]
[1]
[0]
[0]
[MHz]
1
2
1
2