R3765/67G Series Network Analyzer Operation Manual
7.14 Save/Recall
7-148
7.14.7 Purging Stored File
Purges files.
When the file name is defined, the defined file name is displayed on the menu.
The setting and the explanation
Press the [SAVE] to call the save menu. (Refer to section A.4.)
Press the {PURGE FILE} to call the purge file menu.
Purges files menu
{PURGE}
: Purges the file.
{CURSOR }
:
{CURSOR }
:
*
The file of the cursor position can be purged.
7.14.8 Executing Recall
Recalls register or file.
When the register/file name is defined, the defined name is displayed on the menu.
The setting and the explanation
Press the [RECALL] to call the recall menu. (Refer to section A.4.)
Recall menu
•
Recall menu (1 of 4)
{RECALL REG-1}
: Recalls the setting data, calibration data and memory
trace data saved in the register 1.
{RECALL REG-2}
: Recalls the setting data, calibration data and memory
trace data saved in the register 2.
{RECALL REG-3}
: Recalls the setting data, calibration data and memory
trace data saved in the register 3.
{RECALL REG-4}
: Recalls the setting data, calibration data and memory
trace data saved in the register 4.
{RECALL REG-5}
: Recalls the setting data, calibration data and memory
trace data saved in the register 5.
{RECALL POWER OFF}
: The settings are stored automatically when the power of
the R3765/67H series is turned off. When the power is
turned on again, the data is set to the initial state. By
pressing this key, the stored data is read again.
{LOAD FILE}
: Calls the load file menu used to load the all informations
stored in the file. (See step
Note:
Before LOAD FILE is executed, be sure to insert a formatted floppy disk to the
drive.
1
2
3
Shifts the cursor up/down of file list.
1
2
3