![ADVANTEST R3755 Series Скачать руководство пользователя страница 100](http://html.mh-extra.com/html/advantest/r3755-series/r3755-series_operation-manual_2858773100.webp)
7.4
R3760 Specifications
7-10
7.4
R3760 Specifications
This chapter describes about the function of the R3760 and the performance/specification together.
1.
Measurement Function
Measurement channels
4 channels
Measurement parameter
Transmission, Reflection
Measurement format
Log / Linear amplitude, Phase, complex parameter real
and imaginary numbers,
Z, R, X (When measuring impedance conversion)
Y, G, B (When measuring admittance conversion)
2.
Signal Source Section (25
°
C
±
5
°
C, and calibration period: 1 year)
Frequency characteristics
Range
Resolution
Accuracy
300 MHz to 6 GHz
10 kHz
±
50 ppm
Output characteristics
Range
Resolution
Accuracy
Linearity (1 GHz)
Flatness (At 0 dBm Output)
≤
3 GHz:
+0 dBm to -10 dBm
>3 GHz:
-5 dBm to -10 dBm
0.1 dB
±
2.0 dB (0 dBm, 1 GHz)
+0 dBm to -10 dBm
±
1.5 dB
300 MHz to 1 GHz 8.0 dBp-p
1 GHz to 4 GHz
6.0 dBp-p
1 GHz to 6 GHz
8.0 dBp-p
Sweep characteristics
Sweep parameter
Range
Range setting
Sweep type
Sweep time
Measurement points (Segments)
Sweep trigger
Frequency
Frequency sweep same as frequency characteristics
Start / Stop or Center / Span
Any specified segment sweep
(Frequency, Output level, RBW, Point, Settling setting)
Max. 300
µ
s / Point
Max. 1601 points (segments)
Continuous or Single
Output format
Output
Connector
Single (Power splitter embedded)
SMA (Female), 50
Ω