R3267 Series Spectrum Analyzer Operation Manual (Vol.1)
2.2.10 Zooming the Time Domain
2-45
6.
Press SPAN, 5, 0 and MHz.
A frequency span of 50 MHz is set.
7.
Press SWP, Sweep Time AUTO/MNL(MNL), 1, 0, 0 and kHz(ms).
A sweep time of 100 msec is set.
8.
Press COUPLE, RBW AUTO/MNL(MNL), 3 and MHz.
A resolution bandwidth of 3 MHz is set. The burst signal can be identified.
Figure 2-54 Trace of a Burst Signal
9.
Press SPAN and Zero Span.
The frequency span is set to zero span.
10. Press SWP and Trigger Setup.
The Trigger Setup dialog box is displayed.
11. Set Source to VIDEO.
The trigger source is set to VIDEO. The cursor moves to Slope. The trigger level
mark "
→
" is displayed on the left-hand side of the scale.
12. Press Hz(ENTER).
The trigger slope is set to "+" and the cursor moves to Trigger Level.
13. Adjust the trigger level.
Adjust the trigger level to the middle of the burst signal turning the data knob.
A stably triggered display is obtained.
14. Press RETURN.
The Trigger Setup dialog box is removed.