R3267 Series Spectrum Analyzer Operation Manual (Vol.1)
2.2.2 Displaying Spectrums and Operating the Markers
2-21
14. Press 1, 0, MHz(-dBm).
A reference level of -10 dBm is set.
Figure 2-22 Setting Measurement Conditions
Displaying the normal marker on the trace peak
15. Press SRCH.
The normal marker is displayed on the trace peak, and the marker frequency
(approximately 30 MHz) and level (approximately -10 dBm) are displayed in the
marker area.
Figure 2-23 Peak Search
Displaying the delta marker
This measures the frequency difference between a point 3 dB levels down and a point
60 dB levels down from the peak.
16. Press MKR.
The Marker (1) menu used with the marker function is displayed.