R3267 Series Spectrum Analyzer Operation Manual (Vol.1)
5.2.9 GPIB Command Codes
5-41
Table 5-19 UTIL Key (Utility) (3 of 3)
Function
Listener Code
Talker Request
Panel
Key
Remarks
Code
Output Format
IM
measure-
ment
IM measurement mode
ON
IMM ON
IMMON?
0: OFF
UTIL
OFF
IMM OFF
1: ON
Reference wave data read-
out
–––
IMMREF?
Frequency, Level
UTIL
Delta frequency readout
–––
IMMDF?
Delta frequency
Distortion signal data read-
out
–––
IMMRES?
Set number n <DLM>
LL1,LJ1,UL1,UJ1<DLM>
LL2,LJ2,UL2,UJ2<DLM>
....
LLn,LJn,ULn,UJn<DLM>
<DLM>=Delimiter
UTIL
n: Result set number
corresponding to the
degree
LLn: Level difference
in the lower fre-
quency signal
LJn: Pass/Fail judg-
ment result for
the lower fre-
quency signal
0: Pass
1: Fail
-1: Judgment off
ULn: Level difference
for the upper fre-
quency signal
UJn: Pass/Fail judg-
ment result for
the upper fre-
quency signal
Degree setting
IMODR *
IMODR?
Degree
UTIL
Only 3, 5, 7 or 9 can be
specified.
Criteria input
3 rd order
IMLS3 *
IMLS3?
Level
UTIL
5 th order
IMLS5 *
IMLS5?
7 th order
IMLS7 *
IMLS7?
9 th order
IMLS9 *
IMLS9?
Pass/Fail judgment
ON
IMPFC ON
IMPFC?
0 : OFF
UTIL
OFF
IMPFC OFF
1 : ON
Average number
IMAVG *
IMAVG?
Integer (1 to 999)
UTIL
When 1 is specified,
the average function is
turned off.
HI sense mode ON
IMHS ON
IMHS?
0 : OFF
UTIL
OFF
IMHS OFF
1 : ON