5.2.3 GPIB Interface Functions
5-9
(2)
Setting the GPIB Address
Use the following procedure to set the GPIB address for the analyzer:
1.
Press CONFIG and GPIB Address.
The GPIB Address dialog box is displayed.
2.
Use the data knob, the step keys, or the numeric keys to set the GPIB address as
required.
3.
Press ENTR (Hz) to set the address.
(3)
Measurements without displaying characters
When in the remote control mode, measurement speed becomes higher if you turn OFF "Display
ON/OFF."
1.
Press CONFIG and Display ON/OFF(OFF).
OFF is selected, and all indications except for the trace are removed.
5.2.3
GPIB Interface Functions
Code
Description
SH1
Source handshake
AH1
Acceptor handshake
T6
Basic talker, serial polling, listener-specified talker cancel
TE0
Extended talker (not available)
L4
Basic listener function, talker-specified listener cancel
LE0
Extended listener (not available)
SR1
Service request function
RL1
Remote, local, local lockout
PP0
Parallel polling (not available)
DC1
Device clear
DT0
Device trigger (not available)
C0
System controller (not available) (standard)
C1
System controller (option)
C2
IFC transmission, Controller Charging Functions (option)
C3
REN Transmission Function (option)
C4
SRQ Response Function (option)
C12
Interface Message Transmission Function and Control privilege Exchang-
ing Function (option)
E1
Using open-collector bus driver