1.3 Operating Environment
1-4
1.3
Operating Environment
This section describes the environmental conditions and power requirements necessary to use the R3267 Se-
ries.
1.3.1
Environmental Conditions
The R3267 Series should be only be used in an area which satisfies the following conditions:
•
Ambient temperature:
0°C to +50°C (operating temperature)
•
Relative humidity:
85% or less (without condensation)
•
An area free from corrosive gas
•
An area away from direct sunlight
•
A dust-free area
•
An area free from vibrations
•
A low noise area
Although the R3267 Series has been designed to withstand a certain amount of noise riding on the AC
power line, it should be used in an area of low noise. Use a noise cut filter when ambient noise is
unavoidable.
•
An area allowing unobstructed air flow
There is an exhaust cooling fan on the rear panel and exhaust vents on both sides and the bottom
(toward the front) of the R3267 Series. Never block the fan and these vents.
Keep the rear panel 10 centimeters away from the wall. In addition, do not use the R3267 Series
upright turned the rear panel side down. The resulting internal temperature rise will affect measure-
ment accuracy.
Figure 1-1 Operating Environment
Line Filter
Vibration
Direct sunlight
Corrosive
gas
•
Avoid operation in the following areas.
•
Use a noise cut filter when there is a large
amount of noise riding on the power line.
•
Keep the rear panel 10 centimeters
away from the wall
Front
Dust
•
Do not use the R3267 Series upright
turned the rear panel side down.
Front