R3267 Series Spectrum Analyzer Operation Manual (Vol.1)
2.2.14 Pass/Fail Judgments Using the Limit line Function
2-68
16. Press RETURN.
The editor for Limit line 2 is closed and the Limit Line menu is displayed.
Displaying the Limit Line 1 and setting the Pass/Fail criteria
17. Press Limit Line Setup.
The Limit Line Setup dialog box is used to set the Limit Line 1 conditions.
Figure 2-76 Setting Limit Line PASS/FAIL
18. Select ON used with Limit Line 1.
19. Select ‘BELOW the Line’ in Pass Range for the Limit Line 1.
This setting causes data in the area below Limit Line 1 to be considered a pass.
20. Press Limit Line Setup.
The Limit Line Setup dialog box is closed and message PASS is displayed on the
screen after the relationships between Limit Line 1 and the trace data have been
judged as pass.
Figure 2-77 PASS/FAIL Result using Limit Line 1