2.2.14 Pass/Fail Judgments Using the Limit line Function
2-67
11. Continue entering data into Table 2-3, repeating steps 9 and 10.
Figure 2-74 Screen Displayed after Limit Line 1 Data Has Been Entered
12. Press Limit Line 1/2.
The editor is changed from the Limit line 1 mode to the Limit line 2 mode.
13. Press 2, 0 and MHz.
A frequency of 20 MHz is set in the frequency entry used for Limit Line 2. The
cursor moves to the level entry.
14. Press 9, 5 and MHz(-dBm).
A level of -95 dBm is set in the first level entry.
15. Continue entering data into Table 2-4, repeating steps 13 and 14.
Figure 2-75 Screen Displayed after Limit Line 2 Data Has Been Entered