![ADVANTEST R3264 Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/advantest/r3264/r3264_operation-manual_2858762020.webp)
R3267 Series OPT11 3GPP Level Calibration Operation Manual
5.1.3 Performance Verification Test Record Sheet
5-2
5.1.3
Performance Verification Test Record Sheet
The performance verification test record sheet and performance check record sheet is provided at the end
of this chapter.
The test record lists test specification and acceptable limits.
Recommend that make a copy of this table, record the complete test results on the copy, and keep the copy
for calibration test record.
This record could prove invaluable in tracking gradual changes in test result over long period of time.
Table 5-1 Equipment List
No.
Description
Critical Specifications
Recommended
Model
Manufacturer
Notes
1
RF Power
Meter & RF
Power Sensor
Frequency Range: 10 MHz to 3.0 GHz
Power Range: 1
µ
W to 100 mW
VSWR:
≤
1.1
Measurement accuracy: < 0.08 dB
Note: Must be calibrated in advance.
NRVS & NRV-
Z52
Rohde &
Schwarz
PM/PS
2
Signal Genera-
tor
Frequency Range: 10 MHz to 3.0 GHz
Output Level: -15 dBm to +10 dBm
Aging Rate: 1 x 10 exp-6/year
SMP03 equipped
Option B11
Rohde &
Schwarz
SG
3
RF Cable
Impedance: 50
Ω
Connector Type: SMA(m)-SMA(m)
Frequency Range: DC to 3.0 GHz
VSWR:
≤
1.1
Length: 0.7 m approx.
A01002
Advantest
RF_CBL1
4
RF Cable
Impedance: 50
Ω
Connector Type: BNC(m)-BNC(m)
Length: 15 cm approx.
A01036-0150
Advantest
RF_CBL2
5
RF Cable
Impedance: 50
Ω
Connector Type: BNC(m)-BNC(m)
A01037-1500
Advantest
6
Adapter
Type N(f)-SMA(f)
31 N-SMA-50-1/1 HUBER+
SUHNER
Adap_1
7
Adapter
Type N(m)-SMA(f)
HRM-554S
Advantest
Adap_2
8
Adapter
Type N(m)-BNC(f)
JUG-201A-U
Advantest
Adap_3