R3267 Series OPT11 3GPP Level Calibration Operation Manual
2.3 Measurement Sample
2-3
2.3
Measurement Sample
A method of measuring 2-GHz modulation signal Tx power, which uses 3GPP level calibration data, is
explained in this sample.
Resetting the settings
1.
Press SHIFT and CONFIG(PRESET).
The initial settings are read.
Setting the measurement condition
2.
Press FREQ, 2 and GHz.
The center frequency is set to 2 GHz.
Level calibration for the R3267 Series
3.
Connect the A01036-0150 cable that comes with the R3267 Series to the 30 MHz
CAL OUT connector and the INPUT connector.
4.
Press TRANSIENT, STD and Gain Cal.
Gain Cal is performed (this takes approx. 20 seconds).
The level is automatically calibrated for the R3267 Series.
5.
Input a modulation signal to the INPUT connector on the R3267 Series.
6.
Press TRANSIENT, Modulation, Power, Tx Power and Auto Level Set.
The reference level is automatically set.