R3132 Series Spectrum Analyzer Operation Manual
2.2.2 Displaying Spectrums and Operating the Markers
2-19
2.2.2
Displaying Spectrums and Operating the Markers
As an example, the procedure for measuring the difference between calibration signal level and secondary
harmonic signal level is explained below.
Power on
NOTE:
To take accurate measurements, use the spectrum analyzer within the spec-
ified temperature range, and wait at least 30 minutes after turning the power
on before using it. For these practice examples, you do not need to warm up
the spectrum analyzer.
1.
Make sure that the POWER switch on the front panel is in the OFF position.
2.
Connect the power cable provided to the AC power supply connector on the rear
panel.
CAUTION:
To avoid damage, operate the spectrum analyzer within the specified
input voltage and frequency ranges.
3.
Connect the power cable to the outlet.
4.
Turn the POWER switch on.
When the self-test has completed, the startup screen is displayed.
NOTE:
The screen display after the power-on may differ from the one shown here
due to previous settings.
Initialization
This resets the current settings to the factory defaults.
5.
Press SHIFT.
This activates the shift mode so that functions labeled in blue can be used.
6.
Press CONFIG(PRESET).
This resets all spectrum analyzer settings.