R3132 Series Spectrum Analyzer Operation Manual
3.3.18 POWER MEASURE Key (Power Measurement)
3-62
ABS:
Displays an absolute value using the unit selected from
the Units menu under the LEVEL key.
Parameter Setup
Displays the Prmtr Setup menu.
Default
Sets the following settings to default values: the frequency span,
resolution bandwidth, video bandwidth, sweep time, detector,
trace mode, CS/BS Setup,
√
Nyquist Filter Setup and Limit Line
ON/OFF. When Default is enabled, the defaults are automatically
set when the spectrum mask is measured.
Manual
Disables the Default function. As a result, the above settings must
be set manually. The currently set values are used when the spec-
trum mask is measured.
Define
→
Default The currently set values become the new defaults.
Spectrum Mask OFF
Turns the spectrum mask function off.
Spurious Freq
The spurious table information is displayed after the Spuri Freq
menu has been displayed. (The spurious measurement table is not
displayed when there is no table data.)
When making spurious measurements while in frequency mode,
pass/fail judgment is repeated once for each frequency band after
the setup conditions have been selected and a maximum of 10 sig-
nal peaks have been searched for each frequency band. When this
function is used with Average Power measurement, pass/fail
judgments for average powers are performed.
Table No. 1/2/3
Sets Table 1,2 or 3.
1: Sets
Table
1.
2: Sets
Table
2.
3: Sets
Table
3.
Edit Table
Displays the Edit Table menu.
The editor used for the selected table number is displayed. You
can enter the start and stop frequencies, RBW, sweep time and
limit level of the frequency band which is used for pass/fail judg-
ment from this editor.
A maximum of 15 frequency bands can be set.
Table No. 1/2/3
Sets Table 1,2 or 3.
1: Sets
Table
1.
2: Sets
Table
2.
3: Sets
Table
3.