Differential Channel Connections
The differential input channels operate with two signal wires for each channel, and
the voltage difference between both signal wires is measured. On PCIE-1816/
1816H, when all channels are configured to differential input, up to 8 analog channels
are available.
If one side of the signal source is connected to a local ground, the signal source is
ground-referenced. Therefore, the ground of the signal source and the ground of the
card will not be exactly of the same voltage. The difference between the ground volt-
ages forms a common-mode voltage (Vcm ).
To avoid the ground loop noise effect caused by common-mode voltages, you can
connect the signal ground to the Low input. Figure 3-4 shows a differential channel
connection between a ground reference signal source and an input channel on the
PCIE-1816/1816H. With this connection, the PGIA rejects a common-mode voltage
Vcm between the signal source and the PCIE-1816/1816H ground, shown as Vcm in
Figure 3-4.
Figure 3.4 Differential input channel connections