33
The Advanced® Model BR2 Service Manual
9.
Total
or
Direct
LED
does not light when
toggled with the
Total
or
Direct
button on
the front panel.
a. The
Direct
button must be
pressed before pressing the
Direct Timer
button.
b. Pins 3 and 4 of IC2 for a high
after pressing the
Direct Timer
button.
Symptom Check Suggested Action
a. Press the
Direct
button before pressing the
Direct Timer
button.
b. Replace IC2, IC3 or Q4.
10.
Direct Timer
LED
does not light when
the
Direct Timer
but-
ton is pressed.
11.
Direct
LED will light,
but the instrument oth-
erwise appears to
remain in the
Total
mode.
12.
Direct Timer
LED I2
does not flash after
two minutes.
a. Pin 3 of IC3 for a high.
b. Flip-flop Q7-Q8.
a. Replace IC3.
b. Replace Q7, Q8, C20, or C21.
For
Total
, check pin 10 of IC3 for
a high; for
Direct
, check pin 11 of
IC3 for a high.
Replace IC3, J1-J2 ribbon cable, I3, I4, S4,
S5, Q4, Q5, or Q6.
RL1.
Replace RL1.
13.
Direct Timer
time is
not two minutes ±10
seconds.
a. Try resetting R8 (the control
accessible through the round
hole in the right rear of the
instrument cabinet).
b. IC2.
c. R7, R8 and C9.
a. Readjust R8.
b. Replace IC2.
c. Replace R7, R8 or C9.
14. The readout does not
lock when the
Direct
Timer
LED flashes at
the end of a Direct
test.
a. Pin 3 of IC3 for a high.
b. Pin 1 of IC4 for a high.
a. Replace IC3.
b. Replace IC4.
15. Light shining through
the cuvette holder
from the optical lamp
appears too dim or too
bright.
a. X6 and X7 (the LAMP OUT-
PUT terminals on the right rear
side of the Main board) for
+3.35 V.
b. If light is too dim but 3.35 V is
correct, replace I1.
c. Reference diode D4 for +6.9 V.
a. Adjust LAMP ADJ in the right center area
of the Main board.
b. Replace optical lamp I1 (see
Lamp
Adjustment Procedure
).
c. Replace D4.
Содержание Bilirubin Stat-Analyzer BR2
Страница 2: ......
Страница 4: ...4 The Advanced Model BR2 Service Manual Notes...
Страница 5: ...1 Introduction...
Страница 6: ......
Страница 13: ...2 General Overview...
Страница 14: ......
Страница 22: ...22 The Advanced Model BR2 Service Manual Notes...
Страница 23: ...3 Maintenance...
Страница 24: ......
Страница 26: ...Notes...
Страница 27: ...4 Troubleshooting...
Страница 28: ......
Страница 30: ...30 The Advanced Model BR2 Service Manual Notes...
Страница 34: ...34 The Advanced Model BR2 Service Manual Notes...
Страница 35: ...5 Adjustments...
Страница 36: ......
Страница 41: ...6 Replacement Instructions...
Страница 42: ......
Страница 48: ......
Страница 49: ...7 Schematics...
Страница 50: ......
Страница 54: ...Appendix A Symbol Definitions...
Страница 55: ......
Страница 58: ...Appendix B Product Disposal and Recycling...
Страница 59: ......
Страница 61: ...Notes The Advanced Model BR2 Service Manual...
Страница 62: ...Appendix C Service Log...
Страница 63: ......
Страница 65: ...Symptoms Repairs Replacements or Problems or Changes Serviced by The Advanced Model BR2 Service Manual...
Страница 66: ...Index...
Страница 67: ......
Страница 69: ...Notes The Advanced Model BR2 Service Manual...