
OsmoPRO® Multi-Sample Micro-Osmometer User’s Guide
69
8.8 Solenoid
From the
Diagnostics
screen, press
SOLENOID
to enter the
Solenoid Test
screen (Figure
8.11).
Figure 8.11: Solenoid
This screen allows the user to test the functionality of the instruments solenoid.
• Press
IMPACT
to activate the solenoid.
•
Press
EXIT
to return to the
Diagnostics
screen.
8.9 Cleaning
From the
Diagnostics
screen, press
CLEANING
to enter the
Probe Cleaning
screen (Figure
8.12). This screen is where the user can lower the sample probe for cleaning purposes.
Figure 8.12: Cleaning
• The user must remove the turntable, then press
LOWER PROBE
to lower the sample
probe. Once the probe lowers, the user now has easy access to clean the probe.
• After cleaning the probe, press
DONE
to return the probe back to the home position.
• Press
EXIT
to return to the
Diagnostics
screen.