OsmoPRO® Multi-Sample Micro-Osmometer User’s Guide
61
8. Diagnostics
The
Diagnostics
screen (Figure 8.1) allows the user to navigate through the various diagnostic
features designed to test various components of the instrument. Enter diagnostics by pressing
MENU
on the
Main
screen, and then press
DIAGNOSTICS
. From the
Diagnostics
screen, the
user can navigate to the various diagnostic functions discussed in more detail below. To exit
Diagnostics and return to the
Main
screen, press
HOME
.
Figure 8.1: Diagnostics Screen
From the
Diagnostics
screen, press
A/D
to open the
A/D
screen (Figure 8.2). Using this
screen, the user can perform A/D samples. Press
EXIT
to return to the
Diagnostics
screen.
A/D tests are designed to check the performance of the analog to digital converter (A/D),
sample probe, block probe, cooling assembly, and the efficiency of heat transfer between the
components of the cooling assembly. Results are presented as resistance (ohms), temperature
(°C), and power level. Power level is expressed as a percentage and indicates how much power
the thermoelectric cooler uses to reach an operating temperature sensed by the block probe. An
increasing power level, indicates deteriorating performance of the cooling assembly.
Figure 8.2: A/D