15
Immunity/
Compliance
Standard
Compliance
level
a
Electromagnetic environment - guidance
Proximity
Fields from
RF Wireless
Commu-
nications
Equipment
IEC 61000-
4-3
9 - 28
V/m
385
MHz
- 2.45
GHz at
3m
Proximity RF fields generated
by wireless communications
equipment should be at levels
characteristic of a typical location
in a typical commercial or hospital
environment.
Radiate RF
IEC 61000-
4-3
10 V/m
80 MHz
- 2.7
Ghz
80%
AM at 1
kHz
Portable and mobile RF com-
munications equipment should
be used no closer to any part of
the sound processor, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
*Essential performance of the Chorus Sound Processor per
IEC 60601 is defined as auditory stimulation within safe
amplitudes.
RF interference may occur in the vicinity of equipment marked
with the
symbol.
The Chorus Sound Processor is intended for use in electro-
magnetic evironments with controlled RF disturbances. Elec-
tromagnetic interference may be prevented by maintaining
distance between portable or mobile RF transmitters and the
Chorus Sound Processor, as recommended.