2.4.1 Vibration/shock resistance
The vibration/shock resistance tests were performed as follows:
Vibration near machines or conveyor belts
−
DUT: functional device
−
Test standard: EN 60068-2-6
−
Deflection shape: sinusoidal
−
Test axes: X / Y / Z
−
Frequency: 2-9/9-200
−
Change in frequency: + 1 octave/min
−
Deflection: 3 mm
−
Amplitude: 10 m/s²
−
Test duration: 2 h per axis
−
DUT status: DUT electrically in operation
−
Test criterion: Visual inspection after the test and functionality of the DUT during and after the
test
Shock resistance near machines or conveyor belts
−
DUT: functional device
−
Test standard: EN 60068-2-27
−
Shock form: semi-sinusoidal
−
Test axes: +X / +Y / +Z
−
Amplitude: 250 m/s²
−
Duration: 11 ms
−
Test duration: 10 shocks per direction and axis
−
DUT status: DUT electrically in operation
−
Test criterion: Visual inspection after the test and functionality of the DUT during and after the