Chassis Monitoring
Leading
EDGE COMPUTING
3.2.2
Chassis Control and Alarm Threshold Setting
Provides operational and threshold settings for the PXES-2301,
including trigger bus, PXIe link, target temperature, and fan mode,
and threshold settings for DC voltage, temperature, and cooling
fan speed.
Target Temperature
When the Fan switch on the rear panel is set to AUTO, fans run
at different speeds based on the measured temperature.
Target Temp indicates the temperature at which fans run at
100%. Using the default fan curve as an example, fans run at
30% when all temperature readings are less than 25°C, and
begin rampup when any reading exceeds 25°C.
Fans reach 100% speed if any temperature reading exceeds
65°C (i.e. Target Temperature).
Setting parameters are as shown.
Figure 3-3: Target Temperature Parameters
A
Fan
voltage
duty
cycle(%)
0
0
30
100
25
65
Highest temperature (
Ʊ
C) reported by the 3 chassis sensors
B
C
D
E
Содержание PXES-2301
Страница 8: ...viii List of Figures Leading EDGE COMPUTING This page intentionally left blank...
Страница 10: ...x List of Tables Leading EDGE COMPUTING This page intentionally left blank...
Страница 16: ...6 Introduction Leading EDGE COMPUTING Figure 1 2 Left Side View Figure 1 3 Right Side View 319 6 177 8 192 8...
Страница 17: ...Introduction 7 PXES 2301 Figure 1 4 Rear View...
Страница 18: ...8 Introduction Leading EDGE COMPUTING Figure 1 5 Top View...
Страница 19: ...Introduction 9 PXES 2301 Figure 1 6 Underside View...
Страница 26: ...16 Introduction Leading EDGE COMPUTING This page intentionally left blank...
Страница 31: ...Getting Started 21 PXES 2301 Figure 2 1 Cooling Clearances 44 45 44 45 10 5 44 45 44 45...
Страница 33: ...Getting Started 23 PXES 2301 3 Depress the system controller module s latch to release...
Страница 37: ...Getting Started 27 PXES 2301 4 Lift the latch until the module is securely seated in the chassis backplane...
Страница 42: ...32 Getting Started Leading EDGE COMPUTING This page intentionally left blank...
Страница 50: ...40 Chassis Monitoring Leading EDGE COMPUTING This page intentionally left blank...
Страница 54: ...44 Troubleshooting and Maintenance This page intentionally left blank...