97
Click
on the right side of the screen to zero the calibrator.
Click the execution button to start the task.
The upper left of the screen indicates the reading value of the current control pressure, and the upper right
indicates the switch status.
During the test, it can click the graph display button
and the table display button
to switch to graph
display or table display.
The switch action status and action value will be recorded and displayed in the table interface
When the internal pressure module is selected as the standard, the pressure switch test method is based on
slew rate in the control settings and the rate is gradually attenuated following the change of state for the switch.
When using an external pressure module, the control must be done manually.
Click the stop icon at the bottom right of the screen to end the task.
5. Task end
For the contents, refer to Para. 5.2.1 Task end for calibration of the dial pressure gauge.
5.2.6 I/P converter
1. Connection
Connect the calibrator and the I/P converter as shown in Figure 26;
Never apply pressure out of the range of the calibrator.
If the loop power supply is enabled (with maximum load of 50mA), check the load capacity.
2. New task