83
5.1.3 Pressure switch
Select the “pressure switch” icon
in the test type list to start the quick test.
1. Connection
Connect the switch as shown in Figure 25.
2. Start the test
Click the zeroing button
to reset the pressure control module;
Click the start button
to select to enter the execution interface;
Select the graph display button
and the table display button
to switch between the graph display or
table display.
Pressure switch action: When the internal pressure module is used as the reference, it will be automatically run
the test. When the external pressure module is used as the reference, manual pressure generation is required and the
switch data will be recorded.
To stop this test, click the return button .
3. Save the result
Click the save button
to save the test result.
The following items need to be entered: Name and serial number, click OK button
to save the test results.
To exit without saving, press the return button .